Icon Descripon
A tube collision has been detected
A collision of the frontal stand and the lateral stand has been detected (for more in
formaon, see
Intelligent Collision Protecon (page 31))
A lateral stand collision has been detected
A detector collision on the lateral stand has been detected
A tube collision on the lateral stand has been detected
A table collision has been detected
A collision between the stand and the table has been detected (for more informaon, see Intelli‐
gent Collision Protecon (page 31))
A collision between the lateral stand and the table has been detected (for more informaon, see
Intelligent Collision Protecon (page 31))
For more informaon, see Status Area (page 358).
Oponal Conguraons
The system can be customized such that certain BodyGuard sensors are switched o when they are
below the table during APC movements. This oponal funcon is known as BodyGuard O Below Table.
When it is enabled, BodyGuard does not prevent a collision when a part of the paent, such as the
paent’s arm, is below the level of the tabletop. When the stand is posioned at the doctor or nurse
side, the BodyGuard sensor on top of the stand is not switched o, to prevent a collision with the
operator's legs.
About the System Equipment in the Ex
aminaon Room
Azurion Release 1.2 Ins
trucons for Use 30 Philips Healthcare 4522 203 52421