Icon Descripon
A tube collision has been detected
A collision of the frontal stand and the lateral stand has been detected (for more in
formaon, see
Intelligent Collision Protecon (page 31))
A lateral stand collision has been detected
A detector collision on the lateral stand has been detected
A tube collision on the lateral stand has been detected
A table collision has been detected
A collision between the stand and the table has been detected (for more informaon, see Intelli‐
gent Collision Protecon (page 31))
A collision between the lateral stand and the table has been detected (for more informaon, see
Intelligent Collision Protecon (page 31))
For more informaon, see Status Area (page 358).
Oponal Conguraons
The system can be customized such that certain BodyGuard sensors are switched o when they are
below the table during APC movements. This oponal funcon is known as BodyGuard O Below Table.
When it is enabled, BodyGuard does not prevent a collision when a part of the paent, such as the
paents arm, is below the level of the tabletop. When the stand is posioned at the doctor or nurse
side, the BodyGuard sensor on top of the stand is not switched o, to prevent a collision with the
operator's legs.
About the System Equipment in the Ex
aminaon Room
Azurion Release 1.2 Ins
trucons for Use 30 Philips Healthcare 4522 203 52421