14 Maintenance
This pr
oduct requires proper operaon, planned maintenance, and checks that the user must perform
rounely. These tasks are essenal to keep the product operang safely, eecvely, and reliably.
WARNING
Main
tenance of the system by persons without appropriate training, or using unapproved spare
parts, accessories, or detachable parts, may void the manufacturer's warranty. Such maintenance
carries serious risk of personal injury and damage to the system.
Clinical applicaon is not allowed during prevenve maintenance and service.
NOTE The assembly of medical electrical systems and modicaons during the actual service life,
require evaluaon to the requirements of IEC 60601-1.
14.1 Cleaning and Disinfecng
Insucient cleaning of residues that remain on the equipment aer procedures may lead to paent
infecon from polluted parts. Ensure that the system is thoroughly and extensively cleaned aer each
intervenon.
When cleaning and disinfecng the system, follow these general guidelines:
Use sterile covers to prevent polluon or contaminaon of the equipment.
Do not allow liquids to enter the system. This may cause corrosion or electrical damage.
Do not apply cleaning liquid or spray directly onto the system. Always use a cloth dampened with
the cleaning product.
Switch the system o prior to cleaning and disinfecng to avoid electric shock or accidental
acvaon of X-ray. Be aware that even when the system is switched o, live voltages may sll be
present on some interfaces.
The paent straps and ratchet compressor band should be laundered instead of subjected to
surface disinfecon.
Do not use corrosive or abrasive agents or pads.
Some cleaning agents or disinfecon agents may cause discoloraon.
When cleaning scratched or worn painted surfaces, it is to be expected that some addional paint is
removed.
Before cleaning the maress, close the air plug to prevent liquids from entering. Aer cleaning the
maress, open the air plug to allow the maress to expand and contract properly when the paent
is posioned on it.
NOTE You should always comply with local instrucons, regulaons, and guidelines concerning
hygiene.
These cleaning and disin
fecng instrucons only apply to the X-ray system and do not apply to other
equipment in the room. Cleaning instrucons for other equipment are described in the accompanying
documents of the equipment. If cleaning or disinfecng is needed at the interface of third-party
equipment with the X-ray system, dismount the equipment before cleaning or disinfecng. You should
also dismount third-party equipment if you need to clean or disinfect it with agents that are not
compable with the X-ray system.
NOTE Always follow the manufacturer's instrucons for the cleaning agents or disinfectants that
you use.
Maintenance Cleaning and Disin
fecng
Azurion Release 1.2 Ins
trucons for Use 251 Philips Healthcare 4522 203 52421