Item Specicaon
Geometrical ll f
actors Photodiode (opcal ll factor): 63%
Scinllator (X-ray ll factor): 100%
C20/F20
Item Specicaon
Detector 30 x 40 cm (11.6 x 16 inch) (48 cm / 18.9 inch diagonal) 8-
mode a
t detector subsystem
Image format (available at all source-to-image posions) 48 cm (19 inch), rectangular
42 cm (17 inch), square
37 cm (14.4 inch), square
31 cm (13 inch), square
27 cm (10.5 inch), square
22 cm (8 inch), square
19 cm (7 inch), square
15 cm (6 inch), square
Detector rotaon 90 degrees
Detector rotaon me 3 s
Maximum detector rotaon speed 45 degrees/s
Pixel size 154 x 154 µm
Detecve quantum eciency (DQE) 77% at 0 lp/mm
Spaal resoluon properes: MTF:
• 1.0 lp/mm >60%
• 2.0 lp/mm >30%
Nyquist: 3.25 lp/mm
Dynamic range Linear within 2% from 2500 nGy in uoroscopy mode to
50000 nGy in exposure mode (dependent upon operang
mode)
Output digital video 48 cm 1920 x 1448 960 x 742
42 cm 1904 x 1904 952 x 952
37 cm 1688 x 1688 844 x 844
31 cm 1432 x 1432 716 x 716
27 cm 1232 x 1232 616 x 616
22 cm 1016 x 1016 508 x 508
19 cm 864 x 864 432 x 432
15 cm 720 x 720 360 x 360
Geometrical ll factors Photodiode (opcal ll factor): 63%
Scinllator (X-ray ll factor): 100%
16.11 Beam Carriers
F12 S
ystem
The X-ray beam eld is always aligned with the image receptor area and the reference axis is always
perpendicular to the image receptor plane.
Item Specicaon
Rotaon speed 0 - 25 degrees/s (variable)
These speeds ar
e only valid when the stand is in its working
posion, otherwise the maximum speed is 8°/s
Angulaon speed 0 - 25 degrees/s (variable)
These speeds are only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Technical In
formaon Beam Carriers
Azurion Release 1.2 Ins