8 T
o enable secure communicaon, select Use Authencaon.
9 To enable the use of encrypon, select Use Encrypon.
This opon is only available if Use Authencaon is selected.
10 Click Test Connecon.
The result of the test is indicated by an icon.
Test successful
Test failed
If the test fails, more in
formaon is provided.
11 To disable worklist management, select Disabled in the Worklist Management secon.
12 To disable the MPPS manager, select Disabled in the Modality Performed Procedure Step Manager
secon.
13 To undo any changes you have made, click Undo Changes.
14 To save your changes, click Save.
15 To close the System Customizaon window, click Close.
13.7.3 Conguring Remote Systems
You can congure the sengs for other DICOM-compable systems connected to the same hospital
network as the Azurion system.
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the General sengs group, click DICOM Sengs.
3 Click the Remote Systems tab.
A list of remote systems is displayed (DICOM nodes).
4 To view the sengs that are congured for an exisng remote system, select a system in the list.
The Remote System Sengs and Services sengs are displayed, providing informaon about the
selected remote system and the services it supports.
5 To add a new remote system, click Add below the list of remote systems.
A new remote system is added to the list. You can now congure the sengs of the new system.
6 To congure a system's sengs in the Remote System Sengs secon, select the system in the list
and do the following:
a Enter the Name of the remote system.
b Click the Template Type box and select a template.
The template denes the services that available on the remote system. Available services are
indicated with a selected check box in the Services secon.
System Adminis
traon DICOM Sengs
Azurion Release 1.2 Ins
trucons for Use 237 Philips Healthcare 4522 203 52421