14.4 Planned Maintenance Program
To ensure that maintenance is performed at the required intervals, the responsible organizaon should
issue a request to the maintenance organizaon for maintenance to be carried out in accordance with
the Planned Maintenance Program described in this secon.
Planned maintenance may only be carried out by qualied and authorized personnel, and is
comprehensively described in the service documentaon. For more informaon, see Safety (page 18).
Philips Medical Systems provides a full planned maintenance and repair service on both a call basis and
a contract basis. Full details are available from your Philips representave.
A summary of the planned maintenance program appears in the table below. You should always take all
praccal steps to make sure that the Planned Maintenance Program is fully up to date before using the
product with a paent.
Philips Medical Systems will make available, on request, circuit diagrams, component part lists,
descripons, calibraon instrucons, or other informaon that will assist technical support personnel
to repair those parts of the equipment that are designated by Philips Medical Systems as repairable by
technical support personnel.
For a complete list of consumables, contact technical support.
Task Frequency Required Personnel
Check the labels Every 6 months User
Adjust the generators Every 6 months Technician
Adjust the detectors Every 6 months Technician
Perform the level 1 IQ tests Every 6 months Technician
Replace the cooling hoses Every 15 years Technician
Check the coolant levels Every year Technician
Check the x
aon of the ceiling mounted equipment 1 year aer installaon Technician
Check the ceiling rails Every year Technician
Clean the ceiling rails Every year User
Perform mechanical maintenance of the frontal stand Every year Technician
Adjust the frontal stand Every year Technician
Perform mechanical maintenance of the lateral stand Every year Technician
Adjust the lateral stand Every year Technician
Maintenance Planned Maintenance Program
Azurion Release 1.2 Ins
trucons for Use 255 Philips Healthcare 4522 203 52421