5.10 Using an OR Table
Y
ou can use an OR table with the Azurion X-ray system. The level of integraon depends on the OR
table used, and available funcons are described in this secon.
The Azurion X-ray system is compable with OR tables from the following manufacturers:
Maquet
Trumpf
For more informaon, see Compability (page 16).
Tableside Modules
Tableside modules can be mounted on the OR table. During paent transfer to or from the table, the
tableside modules can be parked on a pedestal (opon) in the examinaon room.
Geometry Set-Up and BodyGuard
When using an OR table, the following funcons are not available:
Automac stop at work posions 1 or 2 during motorized longitudinal movement of the stand
BodyGuard O Below Table (oponal funcon)
When using a Maquet operang table, the automac bodyguard override (ABO) funcon is available.
When an object is sensed, this funcon allows stand movements and lateral or longitudinal movements
of the table to connue at a safe speed.
Collision Detecon
WARNING
When mo
ving the table, take care to avoid collisions with the stand.
The intelligent collision pr
otecon (ICP) funcon prevents collisions between the stand and the OR
table base.
When a collision is detected, the following acons are performed:
All table movements are stopped.
A user message is displayed, and an acousc signal sounds.
Normal movement is restored when the collision is resolved.
NOTE When the stand is posioned away from the stored posion, the L-arm starts to rotate when
using the Recall APC funcon, and the C-arm can hit the tabletop during rotaon movement
of the L-arm.
NOTE If the X-ray system is switched o, the collision detecon system will not funcon when table
movement is controlled with the OR table controls.
Accessories
The following accessories can be used with an OR table.
Philips radiolucent arm support (not available with the Trumpf OR table)
Peripheral X-ray lters
Pulse cath arm support
Accessory bracket for ceiling suspended radiaon shield
Ceiling suspended radiaon shield
Cerebral lter