5.10 Using an OR Table
Y
ou can use an OR table with the Azurion X-ray system. The level of integraon depends on the OR
table used, and available funcons are described in this secon.
The Azurion X-ray system is compable with OR tables from the following manufacturers:
• Maquet
• Trumpf
For more informaon, see Compability (page 16).
Tableside Modules
Tableside modules can be mounted on the OR table. During paent transfer to or from the table, the
tableside modules can be parked on a pedestal (opon) in the examinaon room.
Geometry Set-Up and BodyGuard
When using an OR table, the following funcons are not available:
• Automac stop at work posions 1 or 2 during motorized longitudinal movement of the stand
• BodyGuard O Below Table (oponal funcon)
When using a Maquet operang table, the automac bodyguard override (ABO) funcon is available.
When an object is sensed, this funcon allows stand movements and lateral or longitudinal movements
of the table to connue at a safe speed.
Collision Detecon
WARNING
When mo
ving the table, take care to avoid collisions with the stand.
The intelligent collision pr
otecon (ICP) funcon prevents collisions between the stand and the OR
table base.
When a collision is detected, the following acons are performed:
• All table movements are stopped.
• A user message is displayed, and an acousc signal sounds.
Normal movement is restored when the collision is resolved.
NOTE When the stand is posioned away from the stored posion, the L-arm starts to rotate when
using the Recall APC funcon, and the C-arm can hit the tabletop during rotaon movement
of the L-arm.
NOTE If the X-ray system is switched o, the collision detecon system will not funcon when table
movement is controlled with the OR table controls.
Accessories
The following accessories can be used with an OR table.
• Philips radiolucent arm support (not available with the Trumpf OR table)
• Peripheral X-ray lters
• Pulse cath arm support
• Accessory bracket for ceiling suspended radiaon shield
• Ceiling suspended radiaon shield
• Cerebral lter
• Neuro wedge
Preparing a P
aent Study Using an OR Table
Azurion Release 1.2 Ins
trucons for Use 69 Philips Healthcare 4522 203 52421