EasyManua.ls Logo

Philips Azurion - Field Size

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Paent T
ype Weight
Large Adult 70 - 90 kg
Very Large Adult >90 kg
The table above provides guidance for manual paen
t type selecon. You can also select the Automac
paent type. In this case the system automacally selects an appropriate paent type for each study
based on the paent's age, height, and weight, which can be entered while scheduling the paent.
For some applicaons and procedures, the dose sengs are equal for all paent types. In these cases
the automac dose control mechanism manages all depths of irradiated ssue without loss of image
quality, and the paent type selecon has no eect on the reference air kerma (rate). Examples are:
uoroscopy, roadmap, and vascular peripheral. The sengs for the default paent type are used if no
specic X-ray protocols are dened for the selected paent type.
For other applicaons and procedures, the paent type selecon will inuence the reference air kerma.
See the following example for cardiac procedures:
System Paent T
ype Ref. AK (mGy/img)
C12/F12
Cardiac procedure
Le Coronary
15 fps Normal
Neonate 0.041
Infant 0.074
Child, Small Adult 0.117
Default 0.196
Large Adult, Very Large Adult 0.197
C12/F12
Pediatrics procedure
Pediatric
15 fps Normal
Neonate 0.043
Infant 0.074
Child, Small Adult 0.115
Default 0.195
Measurement c
ondions: eld size: 30 cm. All other sengs in accordance with Reference Air Kerma
Measurement Setup (page 311).
For more informaon about the inuence of paent thickness on the air kerma, see Inuence of
Oblique Projecons (page 293).
Field Size
In general, the requested detector dose needs to be larger for smaller eld sizes to compensate for
increasing perceived noise at smaller eld sizes. Therefore, the air kerma and air kerma rate will be
larger for smaller eld sizes.
NOTE Consider zooming uorosc
opy images with appropriate collimaon, instead of using a small
eld size. Digital zooming does not inuence the air kerma.
NOTE Unlike air kerma, dose area product decreases at smaller eld sizes, so using a small eld size
decreases the risk of stochasc eects. For example, for pediatric procedures, a small eld
size may be more appropriate.
For every uoroscopy avor and every exposure X-ray protocol a programmable dose rao per eld size
and per X-ray plane is available. The dose rao indicates per available eld size the percentage detector
dose increase, compared to the detector dose at the largest eld size.
In the examples below, reference air kerma increases approximately proporonally with the dose rao
numbers. The same applies to reference air kerma rate for uoroscopy.
The following example gives the reference air kerma values for a vascular procedure for dierent eld
size values on C12/F12 systems.
Technical In
formaon System Sengs Inuencing the Radiaon Dose
Azurion Release 1.2 Ins
trucons for Use 291 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals