EasyManua.ls Logo

Philips Azurion - 14.2.2 Replacing the Anti-Scatter Grid; 14.3 Replacing Batteries

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 126 P
osioning the detector for removal of the an-scaer grid
2 Move the tabletop just below the detector.
3 Move the spring-loaded locking sliders toward the center of the an-scaer grid.
4 Carefully remove the grid from the detector.
14.2.2 Replacing the An-Scaer Grid
Take care to avoid damaging either the detector or the an-scaer grid during the following procedure.
NOTE Before replacing the an-scaer grid, ensure that it is clean and free from debris.
Figure 127 R
eplacing the an-scaer grid
1 Insert the locang tabs on the an-scaer grid in the corresponding slots of the detector casing.
2 Pull back the locking sliders and push the grid towards the detector unl it is ush with the detector
casing and release the locking sliders.
3 Ensure that the locang tabs are correctly posioned in the detector casing and the grid locking
sliders are correctly engaged.
14.3 Replacing Baeries
For safe operaon, you should replace the baeries in baery-operated equipment at regular intervals.
The baeries of the following items should be replaced regularly:
Maintenance Replacing Baeries
Azurion Release 1.2 Ins
trucons for Use 254 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals