Tomographic Secon Thickness
Cone Beam C
T reconstrucon yields isotropic volumes based on scans made with a square-pixels
detector, therefore the MTF along the rotaonal axis is the same as the MTF in axial slices.
CTDi Dose Measurements
Dose measurements for some typical Cone Beam CT acquision protocols are shown in the table
below.
Phantom Acquision Pr
otocol CTDi (mGy)
CTDI Head XperCT LD 30fps -10s 22
XperCT HQ 30fps -21s 45
CTDI Body Abdomen XperCT Prop LD - 5s 14
Abdomen XperCT Prop HQ -5s 29
Abdomen 3D-RA Prop scan 4s 21
Abdomen 3D-RA Roll scan 8s 21
XperCT Dose Pr
ole
The gure below shows a dose prole for an XperCT acquision protocol.
Technical In
formaon Protecon Against Stray Radiaon