Philips Azurion - 5.10.2 Trumpf Operating Table

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Funcon Universal table top Radiolucent table top
Reset geometry (not available when the stand and table are
lock
ed)
No Yes
Compability with Intervenonal Tools Yes Yes
Alternavely, full c
ontrol of the table is provided on a dedicated Maquet remote control module or
joysck. For details of the Maquet user interface controls, refer to the documentaon supplied with the
Maquet operang table.
NOTE To prevent unintenonal movement of the Maquet table during procedures that require
imaging, it is recommended that you do not use the Maquet controls, and instead use the
two-step approach of the Philips controls: unlock the table and use the movement controls.
NOTE Any movement funcon can be started from either the Xper Geometry module or the Maquet
user interface controls in the examinaon room. However, if a movement funcon is
acvated on each module at the same me, all movement is blocked unl the movement
funcon on both modules is deacvated.
NOTE If movements are blocked on the control module aer changing the table top, rst acvate
the desired movement using the Maquet user interface controls. The control module will then
be enabled again for further movements.
NOTE If the geometry is locked by the control module and the X-ray system is switched o, the
Maquet table is automacally unlocked. Table funcons are sll available using the Maquet
remote control module.
5.10.2 Trumpf Operang Table
The Trumpf operang table consists of the operang table column and tabletop. The tabletop is
available in the following conguraons:
SQ14-XTRA Imaging tabletop
Carbon Floatline Imaging tabletop
Universal tabletop
Only the SQ14-XTRA and Carbon Floatline Imaging tabletops are suitable for X-ray imaging with the X-
ray system.
You can change the tabletop using the Trumpf shule and table docking system.
NOTE When docking the shule to change the tabletop, ensure that the stand is parked. This
provides space for the shule and prevents interference from the X-ray system's collision
prevenon features while docking the tabletop.
NOTE Do not use straps from Philips to secure the paent during movements. Refer to the Trumpf
documentaon for details of how to secure the paent.
Limitaons
When using the T
rumpf operang table, bolus chase and automac posion control of the table are not
available. Automac posion control is available for stand posioning only.
Using Intervenonal Tools with the Trumpf operang table may impose limitaons. Some of the tools,
such as roadmap, require table posion tracking, which is not available on this system. In such cases,
any movement of the table (for example, panning) leads to loss of image registraon.
NOTE The system will not detect a loss of image registraon if the table is panned while using
intervenonal tools.
Preparing a P
aent Study Using an OR Table
Azurion Release 1.2 Ins
trucons for Use 71 Philips Healthcare 4522 203 52421