Philips Azurion - Manual Calibration; Catheter Calibration; Distance Calibration

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The Aut
o calibraon method is automacally selected if the required informaon is available in the
image series.
2 To accept the calibraon factor, click Accept and Connue.
Manual Calibraon
You can perform manual calibraon using one of the following methods:
Catheter
Distance
Sphere
Catheter Calibraon
You perform catheter calibraon by tracing the centerline of a catheter in the image.
You can perform catheter calibraon on either a straight catheter segment or a curved segment, but
you should always use a non-tapered segment. Using a tapered segment for calibraon will result in
incorrect measurement results.
1 Use the navigaon toolbar to review the series and select an image to be used for calibraon.
NOTE You can change the calibraon image at any me by clicking Change in the control panel
and selecng a dierent image.
2 Click the Calibraon task.
3 Click Catheter in the Select calibraon method list.
4 Click Draw in the control panel and do the following:
Click on the centerline of the catheter at the desired start point.
Click again to place a point further along the centerline.
Connue placing points along the centerline, and then double-click at the desired end point.
For addional informaon, click Help in the control panel.
5 To hide or show the contour of the catheter as you work, select or clear Hide in the control panel.
6 To edit a contour, click Edit in the control panel, and do one of the following:
Click along the walls of the catheter in the image and then double-click on the last posion to
complete the contour.
Drag along the walls of the catheter in the image to correct the posion of the contour.
7 When the contours are complete, select the catheter size from the list in the control panel.
If the desired catheter size is not available, you can type it directly in the box.
8 You can delete the calibraon at any me and start over by clicking Delete in the task panel.
9 To accept the calibraon factor, click Accept and Connue in the control panel.
Distance Calibraon
You perform distance calibraon by marking a known distance in the image.
2D Quant
ave Analysis (Opon) QCA / QVA
Azurion Release 1.2 Ins
trucons for Use 162 Philips Healthcare 4522 203 52421