11.1.12 Neuro Wedge
Y
ou can use the neuro wedge to posion the paent's head in the isocenter posion during neuro-
radiology procedures. The neuro wedge should be used with the head support. For more informaon
about the head support, see Head Support (page 193).
1 Slide the tapered end of the neuro wedge under the maress at the head end of the table so that
only the rectangular poron of the wedge is visible.
2 Posion the head support on top the rectangular part of the neuro wedge.
3 Posion the paent so that the paent's head lies comfortably in the head support.
Figure 94 P
osioning the neuro wedge
11.1.13 Ratchet Compressor
The ratchet compressor applies moderate compression to the paent and minimizes paent
movement. This improves visualizaon of internal organs.
1 Posion the unit on the edge of the table and ghten the aachments underneath the unit.
2 Push the release lever down to release the compression band.
3 Pass the compression band over the paent and back under the table, and then put the end of the
band over the compression band roller.
Figure 95 Fing the c
ompression band
4 Turn the ratchet winder clockwise to increase the compression.
Take care to control the amount of compression.
Using Other Equipment Accessories
Azurion Release 1.2 Ins
trucons for Use 194 Philips Healthcare 4522 203 52421