Philips Azurion - 16.2.1 Tube Output Power

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
High Voltage cables
Cover parts
Assembly of X-ray tube and collimator
Item Specicaon
Loading factors corresponding to the maximum specied
ener
gy input to anode in one hour when applied at the
nominal X-ray tube voltage
125 kV, 28 mA (3500 W)
Maximum symmetrical radiaon eld MRC 200+ 0407 ROT-GS 1004: 35 x 35 cm at 1 m distance
MRC 200+ 0508 ROT-GS 1003: 28 x 28 cm at 1 m distance
16.2.1 Tube Output Power
Exposure
Channels with MRC 200+ 0508 ROT-
GS 1003
Channels with MRC 200+ 0407 ROT-
GS 1004
Highest electric power (X-ray tube cur-
r
ent and X-ray tube voltage resulng in
highest electric power)
85 kW (125 kV, 680 mA) 65 kW (125 kV, 520 mA)
Highest electric output power at
100kV, 0.1s (X-ray tube current)
85 kW (850 mA) 65 kW (650 mA)
Maximum X-ray tube voltage and high-
est X-ray tube current at that voltage
125 kV, 680 mA 125 kV, 520 mA
Maximum X-ray tube current and high-
est X-ray tube voltage at that current
1063 mA, 80 kV
The maximum tube current cannot be
reached with the current system con-
guraon.
813 mA, 80 kV
Minimum X-ray tube current and low-
est X-ray tube voltage at that current
10 mA, 40 kV 10 mA, 40 kV
The lowest current me product (load-
ing factors at the lowest current me
product)
0.1 mAs (10 mA, 10 ms) 0.1 mAs (10 mA, 10 ms)
Note: Values ±10%.
Fluoroscopy with Grid Switch
Channels with MRC 200+ 0508 ROT-
GS 1003
Channels with MRC 200+ 0407 ROT-
GS 1004
Highest electric power (X-ray tube cur-
r
ent and X-ray tube voltage resulng in
highest electric power)
25 kW (125 kV, 200 mA) 20 kW (125 kV, 160 mA)
Highest electric output power at
100kV, 0.1s (X-ray tube current)
20 kW (200 mA) 16 kW (160 mA)
Maximum X-ray tube voltage and high-
est X-ray tube current at that voltage
120 kV, 200 mA 120 kV, 160 mA
Minimum X-ray tube voltage and low-
est X-ray tube current at that voltage
40 kV, 1.5 mA 40 kV, 1.5 mA
Maximum X-ray tube current and high-
est X-ray tube voltage at that current
200 mA, 125 kV 160 mA, 125 kV
Minimum X-ray tube current and low-
est X-ray tube voltage at that current
10 mA, 40 kV 10 mA, 40 kV
The lowest current me product (load-
ing factors at the lowest current me
product)
0.007 mAs (2 mA, 3.5 ms) 0.007 mAs (2 mA, 3.5 ms)
Notes:
Values ±10%.
Fluoroscopy is available in pulsed uoroscopy mode only.
Technical In
formaon X-ray System Conguraon
Azurion Release 1.2 Ins
trucons for Use 270 Philips Healthcare 4522 203 52421