Sengs Opons Notes
Processing Format Processed (R
ecommended) Default opon. Image processing is applied to the image
before export.
Unprocessed The image is not processed. Processing parameters are de-
scribed in private DICOM aributes (only IntelliSpace Portal
can handle this correctly).
Select Unprocessed only for export to IntelliSpace Portal or
to a workstaon where measurements are performed on
image data (for example, quantave analysis).
Image Size Do Not Downscale (Recom-
mended for vascular im-
ages)
Default opon
1024x1024
The r
esoluon is limited to 1k
2
. This has no impact on car-
dio images.
512x512
The r
esoluon is limited to 512
2
. The le size is reduced.
Image Quality Normal 8 bits/pixel The le size is reduced.
High 12 bits/pixel (Recom-
mended)
The le size is twice as large as Normal image quality.
d Click Sa
ve to save the new protocol details.
6 To edit an exisng protocol:
a Select the desired protocol in the list.
b Edit the Export Protocol Details.
NOTE If the export protocol has no default desnaon specied, a warning symbol is
displayed in the list.
c Click Save to save the new protocol details.
7 To delete a protocol:
a Select the desired protocol in the list.
b Click Delete the selected export protocol.
c Conrm that you want to delete the protocol.
8 To undo any changes you have made, click Undo Changes.
9 To save your changes, click Save.
10 To close the System Customizaon window, click Close.
13.9 Conguring Automac Data Transfer
You can congure what types of images and data are exported automacally, and what format is used.
For each X-ray protocol, you can specify how you want the system to manage the automac transfer of
image data, by selecng the export protocols to use.
For non-X-ray image data (snapshots, analysis reports, and dose reports), you can select the desnaon
for data based on the data type or the X-ray protocol used to acquire it.
System Adminis
traon Conguring Automac Data Transfer
Azurion Release 1.2 Ins
trucons for Use 241 Philips Healthcare 4522 203 52421