Figure 194 Top view of the paent table
Legend
1 Doctor side 3 Head end
2 Nurse side 4 Foot end
19.1.3 Dose Related Denions
The following denions are used in these Instrucons for Use.
Paent Entrance Reference Point
The paent entrance reference point is an approximaon for the locaon of the paent’s skin (see IEC
60601-1-3:2008, 3.43 and IEC 60601-2-43:2010, 203.5.2.4.5.101d)).
NOTE The distance from focal point to the isocenter can be dierent per type of geometry (see 4.1),
resulng in dierent reference air kerma values under the same circumstances.
NOTE The paent entry reference point may also be known as the intervenonal reference point.
It is located on the central axis of the X-ray beam, 15 cm from the isocenter, towards the focal spot.
Depending on the paent’s size, the table height and the direcon of the X-ray beam, the PERP may be
outside the paent (as in the le gure), may coincide with the skin surface, or may be inside the
paent (as in the right gure).
Figure 195 Paent entrance reference point
Legend
1 X-ray tube 4 Entrance surface
2 Detector 5 Paent entrance reference point
Glossary Denions
Azurion Release 1.2 Ins
trucons for Use 386 Philips Healthcare 4522 203 52421