NOTE Addional ac
cessories may be available from the manufacturer of the OR table. However,
these accessories have not been tested for use with the Azurion X-ray system. Refer to the
informaon provided by the manufacturer.
Parking the Stand
If the oponal ceiling rail extension is installed, you can park the stand so that it is out of the way of the
table. The oponal ceiling rail extension is available at the head end or at the foot end.
5.10.1 Maquet Operang Table
The Maquet operang table consists of the operang table column and tabletop. The tabletop is
available in the following conguraons:
• Universal tabletop: Suitable for a range of surgery applicaons, using a base plate and addional
jointed modules to allow paent posioning.
• Radiolucent tabletop: Suitable for intervenonal procedures and minimally invasive operaons.
NOTE Do not use straps from Philips to secure the paent during movements. Refer to the Maquet
documentaon for details of how to secure the paent.
Startup and Shutdown
Startup and shutdown of the Maquet operang table is managed by the X-ray system. It is not
necessary to power it on or o.
Paent Transfer
During paent transfer, the X-ray system may be switched on or o. If the X-ray system is switched on,
table movement funcons on the control module of the X-ray system are locked.
Paent Orientaon
The Maquet operang table features a blue dot on the table base. When the upper part of the paent's
body is on the same side as the blue dot, the paent orientaon for the Maquet operang table is
Normal (X-ray system: Legs Down). Otherwise, the paent orientaon is Reversed (X-ray system: Legs
Up).
Paent orientaon funcons on each system are linked; when the paent orientaon is changed on the
Maquet operang table, the paent orientaon indicaon of Legs Down or Legs Up on the X-ray system
is updated (the indicaon of Nose Up or Nose Down is not updated). The displayed rotaon angle and
angulaon angle are also updated.
Operaon Modes
A subset of table funcons is available using the control module of the X-ray system.
Funcon Universal table top Radiolucent table top
Basic table funcons (longitudinal, la
teral, height, and cra-
dle movements)
Yes Yes
Isocentric lt No Yes
SyncraTilt No Yes
Automac Posion Control (APC) No Yes
Bolus chase (FDPA) No Yes
Recall APC Yes Yes
Table locking (whole system) Yes Yes
Emergency stop Yes Yes
Preparing a P
aent Study Using an OR Table
Azurion Release 1.2 Ins
trucons for Use 70 Philips Healthcare 4522 203 52421