16.18 Network Data
NOTE Transfer speeds depend on the local situaon (ne
twork load, network devices, and the
external staon).
DICOM Image Interface
Item Specicaon
Maximum Ethernet transfer speed 1 Gbit/s
Transfer speed for images 2 Mbit/s
RIS/CIS DICOM Interface
Item Specicaon
Maximum Ethernet transfer speed 1 Gbit/s
16.19 System Sengs In
uencing the Radiaon Dose
The following secons provide addional informaon about system sengs that inuence the radiaon
dose.
You should also refer to the radiaon guidelines given in Radiaon Safety (page 22) for measures to
reduce paent and sta dose, and to shield stray radiaon.
16.19.1 X-ray Protocol Selecon
The parameters as preset by X-ray protocol selecon are related to each other, and they have been
tuned for an opmal image quality for a specic procedure.
Examples of these parameters are:
• Dose control mode (cine, test shot-lockin, XperCT, etc).
• Timing mode (series, single-shot for cardiac, vascular).
• Dose control curve (for kV, mA, ms, detector-dose).
• Requested detector dose rate – for uoroscopy only.
• Requested detector dose per image – for exposure only.
• Fluoroscopy frame speed (per uoroscopy avor).
• Exposure frame speed (e.g. for cardiac procedures 7.5, 15, or 30 fps).
• Mulphase sengs (e.g. for vascular procedure: the duraon and frame speed per phase).
• Spectral lter (mm Al + mm Cu).
The following examples give the reference air kerma values for typical cardiac, neuro, and vascular X-ray
protocols.
System X-ray Protocol Ref. AK (mGy/img)
C12/F12 Pediatric Pediatrics 15 fps Normal 0.195
Cardiac Pediatrics 15 fps Normal 0.196
Head Cerebral 2 fps Normal 5.051
Head Cerebral 4 fps Normal 5.049
Thorax Lungs 3 fps 2.522
Technical In
formaon Network Data
Azurion Release 1.2 Ins
trucons for Use 289 Philips Healthcare 4522 203 52421