EasyManua.ls Logo

Philips Azurion - Source-to-Image Distance; Table Height; Influence of Oblique Projections

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the wedges reduces the r
adiaon intensity in a user-dened area and improves the image
quality. The wedges also reduce the dose area product and the sta dose.
The amount of radiaon that is reduced by the wedges depends, for example, on the amount of the
image coverage by the wedges.
Source-to-Image Distance
According to the inverse square law, beam intensity increases proporonally with the square of the
distance.
When the source-to-image distance is increased by a factor x, the system increases the skin dose by a
factor x
2
to maintain the requested detector dose.
Hence, the source-to-image distance should be kept to a minimum (for a given source skin distance), so
the requested detector dose is reached with as low as possible skin dose. This implies that the source-
to-image distance should be reduced so that the distance between the paent and the detector is as
small as possible.
Table Height
The table height at a constant source-to-image distance does not inuence the reference air kerma
(rate), and the indicated air kerma (rate) value, as these are only applicable at the paent entrance
reference point.
It does however, inuence the paent skin dose through the inverse square law. For more informaon
on the inverse square law, see Source‐to‐Image Distance (page 293).
To minimize the skin dose (rate), the X-ray source must be as far from the skin as possible.
Inuence of Oblique Projecons
Due to the absorpon of radiaon in human ssue, the X-ray eld strength is reduced by a factor 2,
approximately every 3 cm.
For example, if the paent thickness is 27 cm, the X-ray beam loses intensity within the body by a factor
of 512 (2
(27/3)
). This shows that a thicker paent requires a larger entrance dose than a thin paent, to
obtain the same detector dose.
The same applies to oblique projecons of the X-ray beam since an oblique view generally increases the
perceived paent thickness. This can be seen in the gure below where distance 2 (oblique) is
considerably larger than distance 1.
Figure 132 P
aent thickness
The following example shows that the resulng air kerma is larger for a 30 cm PMMA than for a 20 cm
PMMA paent thickness, when measured at the same system sengs for three typical exposure
procedures.
Technical In
formaon System Sengs Inuencing the Radiaon Dose
Azurion Release 1.2 Ins
trucons for Use 293 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals