Operaon Modes
Y
ou can control the Trumpf table using a dedicated remote control or a control panel on the table
column. For details of the Trumpf user interface controls, refer to the documentaon supplied with the
Trumpf operang table.
When the X-ray system is switched o, it is sll possible to use the Trumpf table.
WARNING
Aer the X
-ray system is switched on, do not move the table unl the X-ray system is fully
operaonal.
Table Control
The Trumpf table can be controlled using the remote control supplied with the table, or using the
emergency control panel on the table base.
NOTE The control module cannot be used for controlling the Trumpf table.
When moving the table towards the tube or towards the a
t detector, the table stops at a distance of
approximately 5 cm from the tube or detector cover.
A collision message appears on the X-ray system display. In this situaon it is not possible to move the
table in any direcon.
NOTE Be aware that the user message disappears aer some me, but t
able movement is sll
inhibited. In this situaon the table can be moved in table override mode. The restricon can
be canceled by moving the stand away from the table.
When the table has been stopped by BodyGuard, the table can be moved in override mode. The
override mode works when the table override joysck is moved downwards and a table movement
buon is pressed simultaneously. Alternavely, you can use the emergency control panel on the table
base which will override movement restricons.
If the movement has been stopped by BodyGuard, the table can also move again when BodyGuard is
no longer acve because the stand has been moved away. The table override joysck does not have to
be used in this situaon.
As long as the override mode is acve, a beep sound is heard and a BodyGuard user message is shown.
CAUTION
It is possible f
or the table to collide with the X-ray system. The table will not stop by itself.
X-ray System Controls
If y
ou are moving the X-ray system towards the Trumpf table and the X-ray system stops because it is
too close (5 cm) to the table, you can move the X-ray system away from the table without overriding.
When BodyGuard is acvated it is possible to move the X-ray system closer to the table at a lower
speed.
During a warm restart procedure and in the X-ray system's fast uoroscopy mode, you can move the
table.
Preparing a P
aent Study Using an OR Table
Azurion Release 1.2 Ins