wiveling the table
1 Clear all objects from the path of the table.
2 Press and hold the direcon buon on the swivel hand switch.
Figure 40 S
wivel hand switch
5.9.7 Using Radiaon Shields
Radiaon shields provide addional protecon against stray radiaon. You can use a table-mounted
radiaon shield and a ceiling-suspended radiaon shield with the system.
The table-mounted and ceiling-suspended radiaon shields are 0.5 mm lead (Pb) equivalent. For
opmal protecon, use the table-mounted and ceiling-suspended radiaon shields together with lead
aprons.
Preparing a P
aent Study Preparing the System
Azurion Release 1.2 Ins
trucons for Use 65 Philips Healthcare 4522 203 52421