11.3.7 8-Meter Cable Assembly Kit (Opon)
Y
ou can use an 8-meter cable assembly kit to connect other equipment to the system and to hospital
power (for example, Philips CX50 ultrasound cart).
NOTE If the 8 meter cable assembly kit is disconnected at the equipment, it is possible that the
cable be le on the oor with live voltage sll present in the connecon plug. This presents a
risk of electric shock if uids make contact with the connectors. To prevent this risk, the
connecon plug must be covered with the rubber cap aached to the connector aer
disconnecng the equipment cable, and the cable must be stored on the wall bracket next to
the wall connecon box.
Figure 116 8-me
ter cable assembly kit
11.3.8 Table Interface Panels
The table interface panels are located under the tabletop and at the rear of the table base.
The table secondary circuit outlet and table base accessory rail provide addional connecvity to the
system. It is possible to route cabling of external equipment through table interface panels. For more
informaon, contact technical support.
These interfaces provide a safe and standardized method for installing third party equipment and do
not impose limitaons on table movements.
Table Secondary Circuit Outlet
The secondary circuit outlet enables you to connect compable medical electrical equipment, or
electrical equipment compliant to IEC basic safety standards.
The secondary circuit outlet power socket is not intended to power medical equipment that has
essenal performance relying on mains power supply. The secondary circuit outlet socket provides up
to 600 VA at 230 V (50/60 Hz).
For more informaon, see Paent Table and Rear Panel Interfaces (page 347).
Using Other Equipment Other Devices
Azurion Release 1.2 Ins
trucons for Use 209 Philips Healthcare 4522 203 52421