EasyManua.ls Logo

Philips Azurion - Page 347

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Surgery Wall Connecon Bo
x
Item Specicaon
IEC or ISO standard IEC / CSA / ANSI / AAMI ES 60601-1
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures The surgery wall connecon box (SWCB) is used with an OR system.
The connector for the pedestal injector is idencal for all injectors (Burndy Met-
alok Bantam 28-pin connector). If this connector is used, a relay is needed within
the SWCB to disconnect the mains supply if the connector is removed. The cur-
rent to be disconnected can go up to 10 A.
The connector for ECG or Physiology equipment is a Burndy Metalok Bantam 23-
pin connector. Because the connector is not safe to touch, the high tension must
be disconnected by means of a relay. The current to be disconnected can go up
to 10 A.
To ground equipment, a grounding cable must be used to connect the equip-
ment to the SWCB.
Equipment Rack
Item Specicaon
IEC or ISO standard ISO11197
Locaon of the equipmen
t Inside the paent environment
Electrical safety measures The equipment rack oers a number of addional power slots that are powered
from the hospital mains directly.
To assure IEC 60601 compliance within the paent environment it is required
that:
Only IEC 60601-1 compliant devices are connected to these power slots.
None of these addional devices has any connecon to equipment from the
Azurion conguraon.
Disregarding these requirements violates the IEC 60601-1 compliance from the
Azurion system.
Intervenonal T
ools Workstaon
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60950
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures Video output 2 of the workstaon shall be connected via Wall Connecon Box to
the video input of a monitor in the examinaon room either directly or via an
oponal Mulvision switch.
Video output 1 of the workstaon shall be connected to the video input of a
slave monitor in the control room either directly or via an oponal Mulswitch.
In case of direct connecon to slave monitor the mains cable of EP navigator
shall be connected to a hospital mains input.
Paent T
able and Rear Panel Interfaces
Item Specicaon
IEC or ISO standard IEC / CSA / ANSI / AAMI ES 60601-1
Locaon of the equipmen
t Inside the Paent environment
Regulatory In
formaon Installaon and Equipment Connecons
Azurion Release 1.2 Ins
trucons for Use 347 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals