Medical DVD Recorder
Item Specicaon
IEC or ISO standard IEC standard
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures Refer to the documentaon provided by the manufacturer of the equipment
Output to TV or Monitor with CVBS Input
Item Specicaon
IEC or ISO standard IEC standard
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures The equipment shall be connected to WVZ out. The TV or monitor shall be a
modern cered TV or monitor with CVBS input, with a cercaon like CE (Low
Voltage Direcve 2006/95/EC) or IEC 60950-1.
Wall Connecon Bo
x
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60950-1
Locaon of the equipmen
t Inside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
It is not permied to connect IEC 60950 equipment, unless a compability exists.
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 40°C / 104°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60950-1
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 35°C / 95°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60601-1
Locaon of the equipmen
t Inside and outside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 35°C / 95°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Regulatory In
formaon Installaon and Equipment Connecons