Medical DVD Recorder
Item Specicaon
IEC or ISO standard IEC standard
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures Refer to the documentaon provided by the manufacturer of the equipment
Output to TV or Monitor with CVBS Input
Item Specicaon
IEC or ISO standard IEC standard
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures The equipment shall be connected to WVZ out. The TV or monitor shall be a
modern cered TV or monitor with CVBS input, with a cercaon like CE (Low
Voltage Direcve 2006/95/EC) or IEC 60950-1.
Wall Connecon Bo
x
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60950-1
Locaon of the equipmen
t Inside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
It is not permied to connect IEC 60950 equipment, unless a compability exists.
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 40°C / 104°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60950-1
Locaon of the equipmen
t Outside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 35°C / 95°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Item Specicaon
IEC or ISO standard IEC / CSA / UL 60601-1
Locaon of the equipmen
t Inside and outside the paent environment
Electrical safety measures Addional equipment and system are powered from the same branch circuit of
hospital mains. Their PE domains are separated.
Addional equipment connected using a wall connecon box shall (can) be con-
nected to WVB-X(ETH), WVB-X(USB), WVB-X(VIDEO).
If the wall connecon box is mounted inside the paent environment, the follow-
ing restricons apply: Ambient temperature shall be below 35°C / 95°F.
Maximum load on 5V output (X4) shall be less than 1A. The wall connecon box
shall be mounted such that its locaon is compliant with Polluon degree 2 (con-
nectors at lower side).
Regulatory In
formaon Installaon and Equipment Connecons
Azurion Release 1.2 Ins
trucons for Use 346 Philips Healthcare 4522 203 52421