EasyManua.ls Logo

Philips Azurion - Page 23

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Select the largest suitable eld siz
e for the current procedure (per X-ray plane).
Use the radiaon disable switch at all mes to prevent accidental exposure to radiaon (except
when the radiaon procedure is in progress).
Shield sensive organs when they are exposed to the beam or are in proximity to it.
Use cauon if the paent has acute skin burns or acute hair loss.
Minimize the duraon of radiaon in uoroscopy and exposure acquision. Modifying sengs like
collimaon, can also be performed while the last image hold image is displayed.
Collimate as much as possible and posion the detector as close as possible to the object.
Keep the paent as far as possible from the X-ray source, using the table height seng.
Keep the focal spot to skin distance as large as possible.
Use dierent X-ray beam projecons, to spread radiaon over the skin.
Avoid oblique projecons, in order to reduce the depth of ssue irradiated.
Consider using uoroscopy instead of exposure acquision.
Clear the primary beam of unnecessary objects. They may cause adverse aects such as
unnecessary paent dose and misinterpretaon of images.
Use only the prescribed air kerma (rate) necessary to perform a procedure.
Release all hand switches and foot switches if the display of live images stops.
Release and depress the hand switch or foot switch again when the requested X-ray does not start
or stop automacally.
Posion the paent and the system as accurately as possible without using radiaon.
Avoid including the table rails in the X-ray image. This can cause unnecessary radiaon load for the
paent.
Sta Safety
Make full use of the system’s radiaon protecon features, devices, accessories, and procedures
that are available to you as the operator. For more informaon, see Using Radiaon
Shields (page 65).
Always wear a lead apron and use badges to monitor the radiaon received.
Stay as far away from the radiated object as possible.
Use cauon if any member of sta has a chronic radiaon injury.
Remove all unnecessary obscuring objects from the primary beam (including the operators hands).
Keep the X-ray source under the table.
Do not aempt to remove, modify, override, or frustrate any safety device on the equipment.
NOTE When door contacts should give a warning for radiaon using the room warning light, the
conguraon of the door contacts should be implemented by the user.
More Informaon
The following table summarizes the eects of the most signicant measures on skin dose rate, air
kerma rate, dose area product, and sta dose.
Measure
Eect on skin
dose r
ate
Eect on
Ref.AK-rate
Eect on DAP-
rate
Eect on sta
dose
Selecng the appropriate X-ray protocol
dose level
+ + + +
Reducing framespeed (by X-ray protocol/
mulphase)
+ + + +
Selecng the largest eld size + + - -
Liming the duraon of uoroscopy/expo-
sure
+ + + +
Applying proper collimaon and wedges 0 0 + +
Increasing the distance from the paent to
the X-ray source (at a constant SID)
+ 0 0 0
Safety Radiaon Sa
fety
Azurion Release 1.2 Ins
trucons for Use 23 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals