Figure 131 T
abletop dimensions: Neuro (top) and Cardio (boom)
Specicaon
Maximum paen
t weight 250 kg
Maximum weight of all accessories (total) 50 kg
Maximum permissible mass (paent plus accessories) 275 kg
Addional allowable force for Cardio Pulmonary Resuscita-
on (CPR)
500 N
Maximum load on the table accessory rail as a result of ac-
cessories (maximum 40 kg on a maximum distance of 0.45
m from the table accessory rail, either horizontal or not)
50 kg
Torque: 184 Nm
Inera: 19 kgm
2
Maximum load on the addional t
able accessory rail as a
result of accessories
10 kg
Torque downward: 40 Nm
Torque upward: 20 Nm
The table can be xed in all posions with use of the brakes -
Addional Accessor
y Weight
It is possible to increase the maximum permissible weight of accessories by reducing the maximum
permissible paent weight under the following condions:
An addional 25 kg above the current maximum weight limit of 50 kg is allowed, increasing the
maximum weight of all accessories to 75 kg.
The maximum paent weight is reduced to 200 kg.
The addional accessory weight should be evenly distributed over the enre length of the two
table accessory rails at the nurse side and doctor side of the tabletop.
Addional accessory weight should not be placed on the table accessory rail at the foot end of the
tabletop.
Technical In
formaon Paent Table
Azurion Release 1.2 Ins
trucons for Use 284 Philips Healthcare 4522 203 52421