Exposure X-ray Protocols: Cardiac (ClarityIQ)
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
X-ray Protocol Paent T
ype Ref. AK (mGy/image)
Le Coronary
15 fps Low
Default 0.018 0.023 0.026 0.029 0.034 0.040 0.046 0.055
Le Coronary
15 fps Medium
Default 0.040 0.050 0.055 0.063 0.073 0.085 0.097 0.116
Le Coronary
15 fps Normal
Default 0.070 0.087 0.095 0.109 0.125 0.145 0.166 0.197
Le Coronary
25 fps Low
Default 0.018 0.023 0.026 0.029 0.034 0.040 0.046 0.055
Rotaonal Scan
Prop Ang0 -4s
Default 0.102 0.127 0.139 0.158 0.181 0.211 0.240 0.285
Infant 0.033 0.040 0.043 0.049 0.055 0.064 0.073 0.087
Neonate 0.020 0.025 0.027 0.032 0.037 0.044 0.051 0.063
Exposure X-ray Protocols: Head
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
X-ray Protocol Paent T
ype Ref. AK (mGy/image)
Cerebral
2 fps Normal
Default 2.279 2.982 3.332 3.937 4.675 5.679 6.716 8.042
Child 1.519 1.863 2.003 2.238 2.517 2.889 3.297 3.822
Infant, Neonate 1.569 1.863 2.003 2.238 2.517 2.889 3.297 3.822
Cerebral
4 fps Normal
Default 2.280 2.982 3.334 3.934 4.669 5.571 6.043 6.776
Child 1.373 1.639 1.768 1.966 2.244 2.630 3.049 3.186
Infant, Neonate 1.373 1.639 1.768 1.966 2.244 2.630 3.049 3.186
3D-RA
Prop 4s
Default 0.115 0.139 0.166 0.167 0.168 0.170 0.172 0.174
Infant, Neonate 0.068 0.085 0.103 0.104 0.105 0.107 0.108 0.110
Exposure X-ray Protocols: Head (ClarityIQ)
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
X-ray Protocol Paent T
ype Ref. AK (mGy/image)
Cerebral
2 fps Low
Default 0.463 0.605 0.677 0.798 0.946 1.149 1.358 1.620
Cerebral
4 fps Low
Default 0.463 0.605 0.676 0.798 0.947 1.149 1.357 1.620
Cerebral
2 fps Normal
Default 0.914 1.195 1.336 1.576 1.871 2.269 2.681 3.204
Cerebral
4 fps Normal
Default 0.914 1.195 1.336 1.578 1.870 2.269 2.684 3.202
3D-RA
Prop Scan 4s
Default 0.115 0.139 0.166 0.167 0.168 0.170 0.172 0.174
Infant, Neonate 0.068 0.085 0.103 0.104 0.105 0.107 0.108 0.110
Technical In
formaon Typical Reference Air Kerma (Rate) Values
Azurion Release 1.2 Ins
trucons for Use 307 Philips Healthcare 4522 203 52421