When you operate the system in emergency access mode, you cannot review other studies. You can
only acquir
e new images and series. You can review the images and series that you acquire while in
emergency access mode but if you end the procedure, you cannot open it again unl you have logged
onto the system.
For more informaon about conguring the system to allow emergency access without logging on, see
Managing Users and System Logon (page 230).
You can start an emergency study without entering any paent details. You can sll nd the study in the
paent database by looking for the me and date of the study contained in the Paent ID.
1 If the system is not switched on, press and hold Power On on the review module unl the indicator
light stops ashing.
2 In the logon screen, click Emergency.
The system is available in emergency access mode. This mode allows you to perform an emergency
procedure, but has reduced funconality.
A study is started immediately using the default ProcedureCard and a menu is displayed allowing
you to select the ProcedureCard for the study.
3 To change the ProcedureCard, do the following:
a Select the paent in the paent list and click Edit.
b Select the appropriate ProcedureCard Group from the drop-down list.
c Select an alternave ProcedureCard.
4 If you are able to, enter any available paent informaon in the Study Details tab.
NOTE You cannot add or change paent details once images have been acquired. If you have
not entered the paent's details before acquiring images, you can add the paent to the
system later when you are logged on and use the Resolve Paent Mix wizard to associate
the acquired series with the paent. For more informaon, see Resolving a Paent
Mix (page 128).
5 To start the study, click Back to Procedure.
6 Perform the necessary procedure.
7 To end the study, do the following:
a Click End Procedure.
A dialog box is displayed with a warning, reminding you that you are in emergency access mode
and that the acquired data will not be accessible if you end the procedure.
b To close the dialog box and connue the study, click Cancel.
c To end the study, click OK.
The study ends and the Add Paent window is displayed, allowing you start another study if
necessary.
8 To start a new study in emergency access mode, click Start Procedure and repeat steps 2 to 6.
9 If all studies are complete and emergency access is no longer needed, click System and select Log
O to exit emergency access mode and return to the logon screen.
Performing Procedures Acquiring Images in an Emergency
Azurion Release 1.2 Ins
trucons for Use 89 Philips Healthcare 4522 203 52421