Before s
tarng the procedure, ensure that the C-arm is posioned at either the nurse or doctor side
and that all objects have been cleared from the table path.
1 Park the lateral stand in the parking posion.
2 Posion the paent on the table.
For more informaon, see Posioning the Paent on the Table (page 52).
3 Posion the peripheral X-ray lters.
For more informaon, see Peripheral X‐ray Filters (page 192).
4 Immobilize the paent's legs.
5 To select the Bolus Chase X-ray protocol, do the following.
a On the touch screen module, tap the X-ray Sengs task.
The X-ray protocols displayed are the ones associated with the currently selected
ProcedureCard.
b Tap Bolus Chase.
If Bolus Chase is not visible in the list of X-ray protocols, tap Other, select Peripherals, and then
select Bolus Chase from the full list of available X-ray protocols.
6 If your system has a rotatable detector, posion the detector in landscape or portrait posion.
7 To allow you to raise the table to its highest posion, raise the detector to the highest possible
posion.
8 Raise the table to the maximum height.
9 Set the eld of view to the maximum size.
10 Center the region of interest at the start posion.
11 Reduce the distance between the paent and the detector to the minimum possible.
12 Lock lateral movements of the table by doing the following:
a On the touch screen module, tap the Table task.
b Tap Lateral to switch the lateral table lock on.
13 Use uoroscopy to conrm that the paent is in the correct posion, by moving the table from the
start posion to the end posion.
14 If necessary, adjust the paent's lateral posion by moving the paent on the tabletop.
NOTE Do not unlock table lateral movements.
15 R
eposion the table longitudinally at the start posion.
16 Switch Injector Coupling on.
For more informaon, see Injector Coupling (page 95).
17 Prepare the injector.
18 Start acquision by pressing and holding the hand switch.
NOTE BodyGuards are disabled during image acquision.
19 When the contrast bolus reaches the boom of the image on the monitor, start moving the
tabletop using the speed controller.
Performing Procedures Bolus Chase
Azurion Release 1.2 Ins
trucons for Use 99 Philips Healthcare 4522 203 52421