Item Specicaon
Angulaon speed 0 - 25 degrees/s (variable)
These speeds ar
e only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Projecon angles Angulaon: 90 degrees caudal to 90 degrees cranial
Rotaon: 185 degrees LAO to 120 degrees RAO
Isocenter to oor 114.0 cm (44.9 inch)
Focal spot to isocenter 81.0 cm (30.9 inch)
Focal spot to detector (source-to-image distance) 89.5 to 119.5 cm (35.2 to 47.1 inch)
Detector movement speed Toward the paent: 10 cm/s
Away from the paent: 15 cm/s
Throat depth 90 cm (35.4 inch)
Motorized stand rotaon 90 degrees le to 90 degrees right at 12 degrees/s
Rotaon scan From 185 degrees RAO (-rotaon) to 120 degrees LAO (+ro-
taon) at a speed of up to 55 degrees/s
F20 System
The X
-ray beam eld is always aligned with the image receptor area and the reference axis is always
perpendicular to the image receptor plane.
Item Specicaon
Rotaon speed 0 - 25 degrees/s (variable)
These speeds ar
e only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Angulaon speed 0 - 25 degrees/s (variable)
These speeds are only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Projecon angles Angulaon: 90 degrees caudal to 90 degrees cranial
Rotaon: 120 degrees LAO to 185 degrees RAO
Isocenter to oor 113.5 cm (44.7 inch)
Focal spot to isocenter 81 cm (31.9 inch)
Focal spot to detector (source-to-image distance) 89.5 to 119.5 cm (35.2 to 47 inch)
Detector movement speed Toward the paent: 10 cm/s
Away from the paent: 15 cm/s
Throat depth 90 cm (35.4 inch)
Motorized stand rotaon 90 degrees le to 90 degrees right at 12 degrees/s
Rotaon scan From 185 degrees RAO (-rotaon) to 120 degrees LAO (+ro-
taon) at a speed of up to 55 degrees/s
C20 System
The X
-ray beam eld is always aligned with the image receptor area and the reference axis is always
perpendicular to the image receptor plane.
Item Specicaon
Rotaon speed 0 - 25 degrees/s (variable)
These speeds ar
e only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Angulaon speed 0 - 25 degrees/s (variable)
These speeds are only valid when the stand is in its working
posion, otherwise the maximum speed is 8 degrees/s
Projecon angles with the C-arm posioned at the head
end of the table
Angulaon: 90 degrees caudal to 90 degrees cranial
Rotaon: 120 degrees LAO to 185 degrees RAO
Projecon angles with the C-arm posioned at the side of
the table
Angulaon: 185 degrees caudal to 120 degrees cranial
Rotaon: 90 degrees LAO to 90 degrees RAO
Technical In
formaon Beam Carriers
Azurion Release 1.2 Ins
trucons for Use 281 Philips Healthcare 4522 203 52421