CALC:LIM:OFFS:AMPL
CALCulate
<Ch>
[:SELected]:LIMit:OFFSet:AMPLitude
<numeric>
CALCulate
<Ch>
[:SELected]:LIMit:OFFSet:AMPLitude?
Description
Sets the value of the limit line offset along Y–axis.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the value of the limit line offset along Y–axis, varies
depending on the data format
Unit
{ dB | ° | s}
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
Analysis > Limit Test > Limit Line Offsets > Response Offset
CALC:LIM:OFFS:MARK
CALCulate
<Ch>
[:SELected]:LIMit:OFFSet:MARKer
Description
Sets the value of the limit line offset along Y–axis to the active
marker value.
(no query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
Analysis > Limit Test > Limit Line Offsets > Marker > Response Ofs