S2VNA and S4VNA SCPI Programming Manual
356
SOUR:POW:PORT:CORR
SOURce
<Ch>
:POWer:PORT
<Pt>
:CORRection[:STATe] {OFF|ON|0|1}
SOURce
<Ch>
:POWer:PORT
<Pt>
:CORRection[:STATe]?
Description
Turns ON/OFF the power correction.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Power Calibration > Correction
SOUR:POW:PORT:CORR:COLL
SOURce
<Ch>
:POWer:PORT
<Pt>
:CORRection:COLLect[:ACQuire]
Description
Measures the power calibration data for the port <Pt> using the
power meter controlled via USB or USB/GPIB. Calculates
calibration coefficients on completion of the measurement, and
turns ON the power correction for the port.
(no query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Equivalent Softkeys
Calibration > Power Calibration > Calibrate

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish