S2VNA and S4VNA SCPI Programming Manual
212
SENS:CORR:COEF:METH:SOLT2
SENSe
<Ch>
:CORRection:COEFficient:METHod:SOLT2
<port1>
,
<port2>
Description
Selects the ports and sets the full 2port calibration type, when the
written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 2 (or 4)
<port2> the second port number from 1 to 2 (or 4)
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COEF:SAVE
Equivalent Softkeys
None
SENS:CORR:COEF:METH:SOLT3
SENSe
<Ch>
:CORRection:COEFficient:METHod:SOLT3
<port1>
,
<port2>
,
<port3>
Description
Selects the ports and sets the full 3port calibration type, when the
written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COEF:SAVE
Equivalent Softkeys
None

Table of Contents