S2VNA and S4VNA SCPI Programming Manual
184
FORM:DATA
FORMat:DATA
<char>
FORMat:DATA?
Description
Sets or reads out the data transfer format, when responding to the
following queries:
CALC:DATA:FDAT?
CALC:DATA:FMEM?
CALC:DATA:SDAT?
CALC:DATA:SMEM?
CALC:FUNC:DATA?
CALC:LIM:DATA?
CALC:LIM:REP?
CALC:LIM:REP:ALL?
CALC:RLIM:DATA?
CALC:RLIM:REP?
SENS:FREQ:DATA?
SENS:SEGM:DATA?
(command/query)
Parameter
<char> Choose from:
ASCii : Character format
REAL : Binary format (IEEE–64 floating point)
REAL32 : Binary format (IEEE–32 floating point)
Query Response
{ASC|REAL|REAL32}
Preset Value
ASC
Notes
The command is not applicable with the TCP/IP Socket protocol.
Related Commands
FORM:BORD
Equivalent Softkeys
None