EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sens:corr:ext:auto:meas

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
284
SENS:CORR:EXT:AUTO:MEAS
SENSe
<Ch>
:CORRection:EXTension:AUTO:MEASure {SHORt|OPEN}
Description
Performs measurement of the standard "SHORT" or "OPEN",
automatically calculates and sets the parameters of the Port
Extension.
The set of ports for which this command is executed is determined
by the SENS:CORR:EXT:AUTO:PORTcommand.
When two consecutive measurements of "SHORT" and "OPEN" are
performed the results of these measurements are averaged.
(command)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
SHORt : Measures "SHORT" standard
OPEN : Measures "OPEN" standard
Equivalent Softkeys
Calibration > Port Extension > Auto Port Extension > Measure Short or
Open

Table of Contents