S2VNA and S4VNA SCPI Programming Manual
365
SOUR:POW:STOP
SOURce
<Ch>
:POWer:STOP
<power>
SOURce
<Ch>
:POWer:STOP?
Description
Sets or reads out the power sweep stop value when the power
sweep type is active.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<power> the power sweep stop value within the power limits of the
analyzer
Unit
dBm (decibels above 1 milliwatt)
Resolution
0.05 dBm
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
Depends on the analyzer
Equivalent Softkeys
Stimulus > Stop
STAT:OPER?
STATus:OPERation[:EVENt]?
Description
Reads out the value of the Operation Status Event Register.
(query only)
Target
Status Reporting System
Query Response
<numeric>
Equivalent Softkeys
None