S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:CKIT:STAN:L3
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3
<numeric>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3?
Sets or reads out the L3 value for the short calibration standard.
(command/query)
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N – the number of the standards in
the calibration kit
<numeric> the L3 value from –1E18 to 1E18
1E–42 H/Hz
3
(Henry/Hertz
3
)
Sets the value of the limit, which is closer to the specified value.
Calibration > Cal Kit > Define STDs > L3 10
–42
H/Hz
3
SENS:CORR:COLL:CKIT:STAN:LAB
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel
<string>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel?
Sets or reads out the label for the calibration standard.
(command/query)
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N – the number of the standards in
the calibration kit
<string>, up to 254 characters
Calibration > Cal Kit > Define STDs > Label