EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sens:corr:coll:ckit:stan:l3; Sens:corr:coll:ckit:stan:lab

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
240
SENS:CORR:COLL:CKIT:STAN:L3
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3
<numeric>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3?
Description
Sets or reads out the L3 value for the short calibration standard.
(command/query)
Target
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N the number of the standards in
the calibration kit
Parameter
<numeric> the L3 value from 1E18 to 1E18
Unit
1E42 H/Hz
3
(Henry/Hertz
3
)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > L3 10
42
H/Hz
3
SENS:CORR:COLL:CKIT:STAN:LAB
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel
<string>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel?
Description
Sets or reads out the label for the calibration standard.
(command/query)
Target
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N the number of the standards in
the calibration kit
Parameter
<string>, up to 254 characters
Query Response
<string>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > Label