S2VNA and S4VNA SCPI Programming Manual
400
TRIG:EXT:SLOP
TRIGger[:SEQuence]:EXTernal:SLOPe
<char>
TRIGger[:SEQuence]:EXTernal:SLOPe?
Description
Sets or reads out the polarity of the external trigger.
(command/query)
Parameter
<char> Choose from:
POSitive : Positive edge
NEGative : Negative edge
Query Response
{POS|NEG}
Preset Value
NEG
Related Commands
TRIG:SOUR
Equivalent Softkeys
Stimulus > Trigger > Ext Trig Polarity > { Negative edge | Positive edge }