Page #141 background imageLoading...
Page #141 background image
S2VNA and S4VNA SCPI Programming Manual
141
CALC:MST:DOM
CALCulate
<Ch>
[:SELected]:MSTatistics:DOMain[:STATe] {OFF|ON|0|1}
CALCulate
<Ch>
[:SELected]:MSTatistics:DOMain[:STATe]?
Description
Selects either partial frequency range or entire frequency range is
used for math statistic calculation. The partial frequency range is
limited by two markers.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
Choose from:
{ON|1} : partial frequency range
{OFF|0} : entire frequency range
Query Response
{0|1}
Preset Value
0
Related Commands
CALC:MST:DOM:STAR
CALC:MST:DOM:STOP
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Range