EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Calc:mark:bwid:thr

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
115
CALC:MARK:BWID:THR
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
:BWIDth:THReshold
<numeric>
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
:BWIDth:THReshold?
Description
Sets the bandwidth definition value..
(command/query)
Target
Marker <Mk> of the active trace of channel <Ch>,
<Ch>={[1]|2|...16}
<Mk>={[1]|2|...16}
Parameter
<numeric> the bandwidth definition value, the range varies depending
on the data format
Unit
{ dB | ° | s }
Query Response
<numeric>
Preset Value
3.0
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Bandwidth Value

Table of Contents