S2VNA and S4VNA SCPI Programming Manual
111
CALC:MARK
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
[:STATe] {OFF|ON|0|1}
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
[:STATe]?
Description
Turns ON/OFF the marker.
Turning ON a marker with the number from 1 to 15 will turn ON all
the markers of smaller numbers. Turning OFF a marker with the
number from 1 to 15 will turn OFF all the markers of greater
numbers (except of the reference marker with number 16). Turning
ON/OFF the reference marker with number 16 does not turn
ON/OFF the markers with the numbers from 1 to 15, but switchs
these markers between relative and absolute measurement mode.
(command/query)
Target
Marker <Mk> of the active trace of channel <Ch>,
<Ch>={[1]|2|...16}
<Mk>={[1]|2|...16}
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Markers > Add Marker | Remove Marker
Markers > Reference Marker