S2VNA and S4VNA SCPI Programming Manual
38
*STB?
*STB?
Description
Reads out the value of the Status Byte Register
(query only)
Target
Status Reporting System
Query Response
<numeric>
Equivalent Softkeys
None
*TRG
*TRG
Description
Generates a trigger signal and initiates a sweep under the
following conditions.
6. Trigger source is set to the BUS (set by the command
TRIG:SOUR BUS), otherwise an error occurs and the
command is ignored.
7. Analyzer must be in the trigger waiting state, otherwise (the
analyzer is in the measurement state or hold state) an error
occurs and the command is ignored.
The command is completed immediately after the generation of
the trigger signal.
(no query)
Target
Analyzer
Related Commands
TRIG:SOUR
Equivalent Softkeys
None