S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:ECAL:ORI:STAT
SENSe:CORRection:COLLect:ECAL:ORIentation:STATe {OFF|ON|0|1}
SENSe:CORRection:COLLect:ECAL:ORIentation:STATe?
Turns ON/OFF the Auto-Orientation function when the AutoCal
Module calibration is executed.
(command/query)
{ON|1} : ON
{OFF|0} : OFF
Calibration > AutoCal > Orientation > Auto-Orientation
SENS:CORR:COLL:ECAL:PATH
SENSe:CORRection:COLLect:ECAL:PATH
<numeric1>
,
<numeric2>
SENSe:CORRection:COLLect:ECAL:PATH?
<numeric1>
Sets or reads out the AutoCal module port number which is
connected to a specified port of Network Analyzer.
(command/query)
<numeric1> Network Analyzer Port Number,
<numeric2> AutoCal Module Port Number:
1- Port A of AutoCal Module
2- Port B of AutoCal Module
Calibration > AutoCal > Orientation > Port 1 | Port 2