SENS:CORR:OFFS:CLE
SENSe
<Ch>
:CORRection:OFFSet:CLEar
Description
Clears the scalar mixer calibration coefficient table.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
None
SENS:CORR:OFFS:COLL:CLE
SENSe
<Ch>
:CORRection:OFFSet:COLLect:CLEar
Description
Clears the calibration measurement data of scalar mixer calibration
when the frequency offset feature is ON.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration >
Cancel