S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:OFFS:COLL:THRU
SENSe
<Ch>
:CORRection:OFFSet:COLLect[:ACQuire]:THRU
<numeric1>
,
<numeric2>
Measures the calibration data of the thru standard of the specified
port when the frequency offset feature is on for scalar mixer
calibration.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<numeric1> Response port number;
<numeric2> Stimulus port number.
If the same port numbers are specified, an error occurs.
The command starts the measurement immediately if the trigger source
for calibration set to the "Internal" by the command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration >
Reflection Port n > Port n @Freq m Thru