S2VNA and S4VNA SCPI Programming Manual
278
SENS:CORR:COLL:THRU:ADD:FULL2:COMP
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL2:COMPlete
<port1>,<port2>
Description
Completes the full 2-port calibration between the specified ports
provided that each port was calibrated using full 1-port calibration:
Measures an arbitrary thru between the ports;
Calculates the error terms Et and El using the unknown thru
algorithm;
Saves the Et and El error terms to the existing calibration
getting the full 2-port calibration from the two 1-port
calbrations.
If the full 2-port calibration already existed between the specified
ports, updates the Et and El error terms.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
Equivalent Softkeys
Calibration > Calibrate > Thru Addition > Comlete 2-Port Calibration