EasyManuals Logo

Copper Mountain Technologies S2 Series User Manual

Copper Mountain Technologies S2 Series
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #239 background imageLoading...
Page #239 background image
S2VNA and S4VNA SCPI Programming Manual
239
SENS:CORR:COLL:CKIT:STAN:L2
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L2
<numeric>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L2?
Description
Sets or reads out the L2 value for the short calibration standard.
(command/query)
Target
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N – the number of the standards in
the calibration kit
Parameter
<numeric> the L2 value from –1E18 to 1E18
Unit
1E–33 H/Hz
2
(Henry/Hertz
2
)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > L2 10
–33
H/Hz
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish