S2VNA and S4VNA SCPI Programming Manual
294
SENS:CORR:INF?
SENSe
<Ch>
:CORRection:INFormation?
<rcvport>
,
<srcport>
Description
Reads out the information string of the calibration acting between the
<rcvport> and <srcport>.
(query only)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<rcvport> the receiver port number from 1 to 2 (or 4)
<srcport> the source port number from 1 to 2 (or 4)
Query Response
<YYYY/MM/DD> <HH:MM:SS>, <Type>, <TypeEx>, <IFBW>,
<Power>, <Temperature>, <CalKit>
<Type>: {RT|RO|RS|F1|OP|F2|F3|F4}
<TypeEx>: {SOLT|SOLR|TRL|COPY}
<CalKit>:
Calibration Kit Label and Description
Equivalent Softkeys
None
SENS:CORR:IMP
SENSe:CORRection:IMPedance[:INPut][:MAGNitude]
<numeric>
SENSe:CORRection:IMPedance[:INPut][:MAGNitude]?
Description
Sets or reads out the system impedance Z0
(command/query)
Parameter
<numeric> the Z0 value from 0.001 to 1000
Unit
Ω (Ohm)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
50
Equivalent Softkeys
Calibration > System Z0