S2VNA and S4VNA SCPI Programming Manual
386
SYST:ERR?
SYSTem:ERRor[:NEXT]?
Description
Reads out the error message of the error occurred when executing
the SCPI commands, from the FIFO (First In First Out) error queue
stored in the Analyzer. The read out error is deleted from the error
queue. The *CLS command clears the error queue. The maximum
size of the queue is 100 messages.
(query only)
Query Response
<numeric>, <string>
Where:
<numeric> error code
<string> error message
If there is no error in the queue, "0, No error" is read out.
Equivalent Softkeys
None
SYST:FREQ:EXT:RFP:POW
SYSTem:FREQuency:EXTender:RFPort:POWer
<numeric>
SYSTem:FREQuency:EXTender:RFPort:POWer?
Description
Sets or reads out the RF Port Power when analyzer is configured to
work with a frequency extender.
(command/query)
Parameter
<numeric> the power value.
Unit
dBm (decibel relative to 1 milliwatt)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
System > Misc Setup > Frequency Extender > RF Port Power