Sets or reads out the approximate delay value of an adapter in the
adapter removal/insertion function. This value is used to eliminate
the uncertainty of ± 180 ° when calculating the phase response of
the adapter.
The sign of the value depends on the type of the removal /
insertion function. The value must be negative for the adapter
removal function and must be positive for the adapter insertion
function.
If this value is set to zero, the analyzer uses an algorithm to
automatically determine the delay of the adapter. In most cases
setting this value to zero is enough. Setting this value to non zero
is required when:
(command/query)