EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sour:pow:span

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
363
SOUR:POW:SPAN
SOURce
<Ch>
:POWer:SPAN
<power>
SOURce
<Ch>
:POWer:SPAN?
Description
Sets or reads out the power span when the power sweep type is
active.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<power> the power sweep span value from 0 to maximum limit of the
analyzer
Unit
dBm (decibels above 1 milliwatt)
Resolution
0.05 dBm
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
Depends on the analyzer
Equivalent Softkeys
Stimulus > Span